FRT CFM - Sensor for Confocal Microscopy


The confocal measuring principle used in the FRT CFM sensor is an established, non-contact, measuring method for very fast, high-resolution 3D analyses of small structures in research and production like MEMS, microlenses, defects and more. Beside the investigation of microstructures, the sensor is also used for roughness measurements and the capture and analysis of 3D topography. Due to its large-area, confocal measuring principle, meaningful results are generated in less than 10 seconds. The field of view can be extended through the integrated stitching function.

Eigenschaften

  • Optical, non-contact, non-destructive measurement
  • Very fast, large-area measurement
  • Nanometer vertical resolution, sub-micrometer lateral resolution
  • Extendible field of view through stitching
  • Durable, low-maintenance LED light source



Typische Anwendungen

  • Structure measurement
  • Roughness measurement
  • 3D topography measurement

 

FRT CFM - Sensor for Confocal Microscopy 


         Download Options
Technische Daten

Objective 1) 10X 20X 50X 100X
Working distance 1 mm
Resolution z 10 nm 3 nm 2 nm 1 nm
Resolution x.y 2.31 µm 1.16 µm 0.463 µm / 0.42 µm 2) 0.231 µm / 0.37 µm 2)
Field of view 1780 µm x 1335 µm 890 µm x 655 µm 356 µm x 267 µm 178 µm x 134 µm
Numerical aperture 0.5 0.75 0.8 0.9
Measuring angle 90° ± 30° 90° ± 48° 90° ± 53° 90° ± 64°
1) Further objectives available (e.g. long-distance)
2) Geometric / optical
Sensor
Camera 768 x 582 Pixel
Travel range z 400 µm
Typical measuring time 5 - 10 sec.
Light source LED, 505 nm, 5 W
Operating temperature 25° C ± 2° C
Dimensions (W x H x D) 140 mm x 450 mm x 140 mm
 
Scope of Delivery
Measuring head, one objective of choice, operating manual